Pengaruh Variasi Waktu Penundaan Pemeriksaan Terhadap Kadar Elektrolit (Na, K, Ca) pada Sampel Serum di RSUD dr. Doris Sylvanus

Detail Cantuman

CD-ROM

Pengaruh Variasi Waktu Penundaan Pemeriksaan Terhadap Kadar Elektrolit (Na, K, Ca) pada Sampel Serum di RSUD dr. Doris Sylvanus

XML

ABSTRAK

Pengaruh Variasi Waktu Penundaan Pemeriksaan Terhadap Kadar
Elektrolit (Na, K, Ca) pada Sampel Serum di RSUD dr. Doris Sylvanus

Peneliti: Prisma Selvia

Pembimbing: Yayuk Kustiningsih, Anny Thuraidah

Pemeriksaan elektrolit darah merupakan salah satu pemeriksaan
laboratorium klinik yang sangat dipengaruhi oleh tahapan pra analitik. Penundaan
pemeriksaan adalah salah satu faktor yang dapat mempengaruhi pemeriksaan
elektrolit. Tujuan penelitian ini untuk mengetahui pengaruh variasi waktu
penundaan pemeriksaan terhadap kadar elektrolit (Na, K, Ca) pada sampel serum
di RSUD dr. Doris Sylvanus. Metode penelitian ini menggunakan metode
experiment dan menggunakan rancangan post test only with control. Subjek
penelitian ini adalah pasien medical check up berjenis kelamin laki-laki yang
dinyatakan sehat dan tidak mengidap suatu penyakit. Pemeriksaan sampel
menggunakan elektrolit Analyzer GEM 3500 Premier dengan metode ISE (Ion
Selective Elektrode). Analisis data menggunakan program komputer dengan Uji
Repeated Anova. Hasil penelitian menunjukkan rata-rata kadar natrium 0 jam, 1
jam, 2 jam dan 3 jam berturut-turut adalah 139 mmol/L, 139 mmol/L, 141 mmol/L

dan142 mmol/L. Rata-rata kadar kalium 0 jam, 1 jam, 2 jam dan 3 jam berturut-
turut adalah 4,2 mmol/L, 4,2 mmol/L, 4,3 mmol/L, dan 4,4 mmol/L. Rata-rata

kadar kalsium 0 jam, 1 jam, 2 jam dan 3 jam berturut-turut adalah 1,17 mmo/L,
1,15 mmol/L, 1,13 mmol/L, dan 1,11 mmol/L. Uji Repeated Anova menunjukkan
hasil natrium dengan signifikansi 0,000 (< 0,05), kalium dengan signifikansi 0,009
(< 0,05), kalsium dengan signifikansi 0,000 (< 0,05) artinya ada perbedaan yang
bermakna dari variasi waktu penundaan pemeriksaan terhadap kadar natrium,
kalium dan kalsium dengan variasi waktu 0 jam, 1 jam, 2 jam dan 3 jam.
Kesimpulan ada pengaruh variasi waktu penundaan pemeriksaan terhadap kadar
natrium, kalium dan kalsium pada sampel serum dengan variasi waktu 0 jam, 1 jam,
2 jam dan 3 jam. Saran penelitian dapat dilakukan penelitian lebih lanjut terkait
pengaruh suhu penyimpanan sampel terhadap kadar elektrolit (Na, K, Ca).

Kata kunci: Kadar elektrolit (Na, K, Ca), Waktu Penundaan Pemeriksaan

ABSTRACT

Effect of Time Variation of Electrolyte Examination Delay (Na, K, Ca)

on Serum Samples at RSUD dr. Doris Sylvanus

Author: Prisma Selvia

Advisor: Yayuk Kustiningsih, Anny Thuraidah

Blood electrolyte examination is one of the clinical laboratory tests that is strongly
influenced by the pre-analytical stage. Examination delay is one of the factors that
can affect electrolyte examination. Every 30 minutes after the blood draw, a delay
in the examination of the blood sample allows the electrolyte levels in the tube to
decrease. The purpose of this study was to determine the effect of variations in test
delay time on electrolyte levels (Na, K, Ca) in serum samples at RSUD dr. Doris
Sylvanus. This research method uses an experimental method and uses a post test
only design with control with purposive sampling techniques. The subjects of this
study were male medical check-up patients who were declared healthy and did not
suffer from a disease. Sample examination using the GEM 3500 Premier Analyzer
electrolyte. Data analysis using a computer program with the Repeated Anova Test.
The results showed that the average sodium rate of 0 hours, 1 hour, 2 hours and 3
consecutive hours was 139 mmol/L, 139 mmol/L, 141 mmol/L and 142 mmol/L.
The average potassium rate of 0 hours, 1 hour, 2 hours and 3 consecutive hours is
4,2 mmol/L, 4,2 mmol/L, 4,3 mmol/L, and 4.4 mmol/L. The average calcium rate
of 0 hours, 1 hour, 2 hours and 3 consecutive hours are 1.17 mmo/L, 1.15 mmol/L,
1.13 mmol/L, and 1.11 mmol/L. Repeated Anova test shows sodium results with
sig. 0.000 (< 0.05), potassium with sig. 0.009 (< 0.05), calcium with sig. 0.000 (<
0.05) means that there is a significant difference from the variation in the delay time
of examination of sodium, potassium and calcium levels with time variations of 0
hours, 1 hour, 2 hours and 3 hours. Conclusion there is an influence of variations in
test delay time on sodium, potassium and calcium levels in serum samples with time
variations of 0 hours, 1 hour, 2 hours and 3 hours. Research suggestions can be
carried out further research related to the effect of sample storage temperature on
electrolyte levels (Na, K, Ca).

Keywords: Electrolyte Levels (Sodium, Potassium, Calcium), Examination Delay Time


Detail Information

Item Type
Penulis
Prisma Selvia - Personal Name
Yayuk Kustiningsih, M.Kes - Personal Name
Dra. Anny Thuraidah, Apt.,MS. - Personal Name
Student ID
Dosen Pembimbing
Penguji
Kode Prodi PDDIKTI
Edisi
Departement
Kontributor
Bahasa
Indonesia
Penerbit Jurusan Analis Kesehatan : Banjarbaru.,
Edisi
Subyek
No Panggil
SKP AJ 024J TLM/ 2023
Copyright
Doi

Lampiran Berkas

LOADING LIST...



Informasi


DETAIL CANTUMAN


Kembali ke sebelumnya  XML Detail